Swivel Bar Replacement; Base Adjustment - Invacare 9700 Installation And Operating Instructions Manual

Manual / electric portable patient lift and slings
Table of Contents

Advertisement

CARE AND MAINTENANCE OF YOUR PATIENT LIFT
Swivel Bar Replacement (FIGURE 3)
WARNING
After the first year of use, the hooks of the
swivel bar and the mounting brackets of the
boom should be inspected every three (3)
months to determine the extent of wear. If these
parts become worn, replacement must be
made.
1. The swivel bar comes attached to the boom with
covers, bolt, nut, washers, plastic bushing and
spacer.
NOTE: Washers should be placed against both
sides of the boom mounting bracket to
eliminate damage.
2. Plastic covers are placed over the bolt and nut.
NOTE: The swivel bar has three (3) hooks on
both ends to attach the sling that supports the
patient during lift.
3. Remove existing hardware and replace swivel
bar.
CAUTION
DO NOT overtighten mounting hardware. This will
damage mounting brackets.
FIGURE 3 - SWIVEL BAR REPLACEMENT
CARE AND MAINTENANCE OF YOUR PATIENT LIFT
Base Adjustment (FIGURE 4)
The base adjustment should not require any
attention other than:
1. Check the squareness of the legs when in the
CLOSED position.
2. Position the stops so that linkage rods can be
adjusted to obtain 90
3. Adjust the linkage rods until 90
achieved.
4. Reposition the stops after base legs are adjusted
to maintain 90
Stop
FIGURE 4 - BASE ADJUSTMENT
35
O
alignment.
O
alignment is
O
alignment.
Linkage
Rods
Stop
C
A
R
E
-
-
M
A
I
N
T
E
N
A
N
C
E