Chapter 5 Service
5.2 Cover gasket replacement
Inspect the cover gasket periodically, e.g. monthly, for integrity, no damage and replace as needed.
1. Pull out the work tray and take it out.
2. Remove cover gasket from the notch.
3. Insert a new gasket in the notch and take care that the gasket lip is outside.
Warning Contamination hazard!
Damaged or improperly mounted membranes can contaminate the user or the environment!
4. Insert the work tray and push into the interior.
Cell Locker 50155007_C 27.05.21