Performance Verification

System Operation Checks

System Operation Checks
Use the following procedures and the standard section in this kit to verify the operation of your calibrated
system.
NOTE
If you are using a different analyzer, refer to its documentation for specific measurement
instructions.

8510 System Operation Check

NOTE
The 8510 network analyzer is no longer being sold or supported by Keysight. However, you
can download the 8510 class assignments and standard definitions from Keysight's
Calibration Kit Definitions Web page at
1. Short the standard section and perform an S
reflection of the shorted waveguide section does not exceed the limits listed in the tables on
If the ripple is less than the acceptable value, then the combination of directivity and source match
errors of the calibrated system is within the factory specified values.
2. Check the S
and S
11
22
the directivity error, but, because of the thru connection to port 2, this measurement also contains
errors contributed by other terms, including load match, the S
transmission tracking. The actual or absolute return loss of the standard section should not exceed the
acceptable values listed on
3. Make an S
and an S
21
verify that the combination of source match error and load match error do not cause excess ripple on
the trace. The trace ripple should not exceed the acceptable values listed in the tables on
4. Record the results of each test in
The S
and S
of a Shorted Standard Section
11
22
1. Set the averaging factor to 512.
2. Perform a TRL calibration.
3. Turn on the calibration.
4. Connect the standard section to port 1.
5. Connect the short to terminate the standard section.
6. Connect the load to port 2.
5-6
11
of a thru measurement. As in the previous step, the results of this check contain
page
5-11.
measurement. Ideally, the standard section is low loss. These measurements
12
Table 5-1
through 5-3.
www.na.tm.keysight.com/pna/caldefs/stddefs.html
and an S
measurement to verify that the ripple on the
22
and S
11
22
page
of the standard sections, and
page
5-11.
X, P, and K 11644A
5-11.