Technical Data; Guaranteed Values; Power Supply; Analog Measurement Inputs - Omicron DANEO 400 User Manual

Table of Contents

Advertisement

8

Technical data

8.1

Guaranteed values

The values are valid for the period of one year after factory calibration, within 23 °C ±5 °C at
nominal value and after a warm-up time greater than 25 min.
The given input/output accuracy values relate to the range limit value (% of range limit value).
8.2

Power supply

Main power supply
Connection
Voltage, single phase
nominal voltage
operational range
Power fuse
Nominal current
Frequency
nominal frequency
operational range
Overvoltage category
8.3

Analog measurement inputs

Analog inputs: ANALOG INPUT 1...10
Number of analog measurement inputs
Measurement ranges
(RMS value of the sinusoidal shaped
input signal)
Crest factor
Sampling frequency
Input impedance
Measurement category
Temperature drift
Connector according to IEC 60320
100 ... 240 V AC
85 ... 264 V AC
T 12.5 AH 250 V (5 x 20 mm)
"Schurter", order number 0001.2515
10 A
50/60 Hz
45 ... 65 Hz
II
10
10 mV, 100 mV, 1 V, 10 V, 100 V, 600 V
1.75
10 kHz
40 kHz
(1 MΩ ±2%) || (170 pF ±50 pF)
CAT II / 600 V
CAT III / 300 V
CAT IV / 150 V
<25 ppm/K
Technical data
63

Advertisement

Table of Contents
loading

Table of Contents