Mechanical stage S mot. CAN
Measuring stage S mot. CAN
3.2.3
Mounting the mechanical stage /
measuring stage S mot. CAN on the
transillumination base 300
•
If necessary, remove the glass plate (Fig. 7/1)
from the transillumination base 300 (Fig. 7/2) by
pressing on the rear edge to tilt it.
•
Insert stage adapter 84+120 (Fig. 7/3) into the
transillumination base 300 and screw it hand-
tight with the 4 captive screws.
•
Set the mechanical / measuring stage (Fig. 7/5)
on the transillumination base 300 (stage drive
pointing to the left), align to the rear edge and
secure with the three captive screws (Fig. 7/4) of
the stage.
•
Insert the glass plate (Fig. 7/6) into the stage.
The procedure for inserting the plate
into and removing from the stage is
described on page 15.
The red dots on the table (Fig. 7/5) and
metal plate (Fig. 7/6) are to ensure
reproducibility of the alignment of the
metal plate and stage if the former
needs to be detached.
08/2012
SET-UP
Mounting the mechanical / ...
Fig. 7
435465-9000-701
mechanical /
Mounting the
measuring stage mot.
transillumination base 300
Carl Zeiss
on
17