Carl Zeiss
Fig. 5
Mounting the adapter ring
Fig. 6
Mounting the
measuring stage mot.
stand base 450
16
Mounting the mechanical / ...
mechanical /
on
435465-9000-701
SET-UP
3.2.2
Mounting the mechanical /
measuring stage S mot. CAN on
stand base 450 for epi-illumination
If necessary, remove the metal plate (Fig. 5/1)
from insert plate (Fig. 5/3) by pressing and
tilting the rear edge.
If stand base 450 (Fig. 5/5) is not provided with an
adapter ring (Fig. 5/4) this must be mounted first:
− Remove the insert plate (Fig. 5/3). To do this,
remove the three screws (Fig. 5/2) and keep
them for possible later use.
− Insert the adapter ring (Fig. 5/4) into the
stand base 450 and secure with the three
captive screws (Fig. 5/6).
− Detach the insert plate (Fig. 5/3) from the
stand base 450 (Fig. 5/5).
Set the mechanical / measuring stage (Fig. 6/3)
on the insert plate (Fig. 6/1) (stage drive
pointing to the left), align to the rear edge and
secure with the three captive screws (Fig. 6/2) of
the stage on the transillumination top.
Insert the metal plate (Fig. 6/4) into the stage.
The procedure for inserting the plate
into and removing from the stage is
described on page 15.
The red dots on the stage (Fig. 6/3)
and metal plate (Fig. 6/4) are to ensure
reproducibility of the alignment of the
metal plate and stage if the former
needs to be detached.
Mechanical stage S mot. CAN
Measuring stage S mot. CAN
08/2012