Technical Data - Zeiss Mechanical stage S mot. Operating Manual

For axio zoom and stereo discovery
Table of Contents

Advertisement

Mechanical stage S mot. CAN
DESCRIPTION
Measuring stage S mot. CAN

Technical data

Carl Zeiss
2.5
Technical data
Dimensions
Mechanical stage / measuring stage S mot. CAN (length x width x height) ............ 532 mm x 407 mm x 98 mm
Weight
Mechanical stage / measuring stage S mot. CAN ............................................................................. 7.76 kg
P ayload .................................................................................................................................... max. 3 kg
Power supply
Mechanical stage / measuring stage S mot. CAN ................................................ via
CAN extension cable HD15
.......................................................................................................... from
Controller EMS 3 or focus motor
Travel speed
Via Joystick .......................................................................................................................... max. 17 mm/s
Memory travels ................................................................................................................. max. 100 mm/s
Ambient conditions
Storage (in packaging)
Permissible ambient temperature+5 to +40 °C
Permissible relative humidity max. 75 % at +35 °C (non-condensing)
Transport (in packaging):
Permissible ambient temperature ........................................................................................... -40 to +70 °C
Operation
Permissible ambient temperature .......................................................................................... +10 to +40 °C
Permissible relative humidity max. 75%
Air pressure ............................................................................................................... 800 hPa to 1060 hPa
Degree of pollution2
EMS 3 controller
Operational area ................................................................................................................... Closed rooms
Protection class I
Ingress protection rating .................................................................................................................... IP 20
Electrical safety conforms to DIN EN 61010-1 (IEC 61010-1)
and CSA and UL regulations
Overvoltage category ............................................................................................................................. II
RFI suppression ......................................................................................... conforming to EN 55011 Class A
Noise immunity ........................................................................................... conforming to DIN EN 61326-1
Mains voltage range .............................................................................................. 100 to 240 V AC ± 10%
Mains frequency ..................................................................................................................... 50 to 60 Hz
Power consumption .............................................................................................................. max. 180 VA
Fuses .............................................................................................................................. 2xT 4 A/H 205 V
08/2012
435465-9000-701
13

Advertisement

Table of Contents
loading

This manual is also suitable for:

Measuring stage s mot.

Table of Contents