Mounting The Mechanical / Measuring Stage S Mot. Can On The Transillumination Top Dl 450 Mot - Zeiss Mechanical stage S mot. Operating Manual

For axio zoom and stereo discovery
Table of Contents

Advertisement

Carl Zeiss
Fig. 8
Mounting the
measuring stage mot.
transillumination top DL 450 mot.
18
Mounting the mechanical / ...
mechanical /
on
435465-9000-701
SET-UP
3.2.4
Mounting the mechanical /
measuring stage S mot. CAN on the
transillumination top DL 450 mot.
The procedure for mounting trans-
illumination top DL 450 mot. on base
plate 450 is described in the separate
operating
illumination top.
Remove the glass plate (Fig. 8/1) from insert
plate (Fig. 8/3) by pressing and tilting the rear
edge.
Insert assembly aid d = 120 mm (Fig. 8/2) to
prevent the screws and small parts from being
lost in the transillumination top (Fig. 8/4).
Remove the three short hexagon socket screws
(Fig. 8/5) on the insert plate and keep for
possible later use.
Remove assembly aid d = 120 mm (Fig. 8/2)
again.
Set the mechanical / measuring stage (Fig. 8/7)
on the insert plate (stage drive pointing to the
left),
align
transillumination top (Fig. 8/4) and secure on
the transillumination top with the three captive
screws (Fig. 8/6) of the stage.
Insert the glass plate (Fig. 8/8) into the stage.
The procedure for inserting the plate
into and removing from the stage is
described on page 15.
The red dots on the table (Fig. 8/7) and
glass plate (Fig. 8/8) are to ensure
reproducibility of the alignment of the
metal plate and stage if the former
needs to be detached.
Mechanical stage S mot. CAN
Measuring stage S mot. CAN
manual
of
the
to
the
rear
edge
trans-
of
the
08/2012