3.2.2 Setting Auto Sipper

(2) Select a measurement item. For the details of each
measurement item, refer to the section "4.1 What the Product
Can Do" of "INSTRUCTION MANUAL FOR MODEL UH5300
SPECTROPHOTOMETER".
Fig. 3-4 Measurement Item Selection Screen
Select Auto Sipper from the tabs, and set Sipping Time, Delay Time,
and Air Purge. For setting other measurement conditions, refer to the
chapter "4 First-Time Measurement" of "INSTRUCTION MANUAL
FOR MODEL UH5300 SPECTROPHOTOMETER".
Fig. 3-5 Auto Sipper Setting Screen
Sipping Time: This is the processing time for sipping sample
caused by pushing the sipping lever. The time is
settable within the range of 0 - 60.0 seconds at
0.1 second intervals. When the viscosity of sample is
equal to that of pure water, approx. 1.5 mL can be
sipped in 5 seconds. To measure a sample with a
viscosity equivalent to pure water at the room
temperature of 25 °C, it is recommended for sipping
for 5 seconds.
3 - 3
3.2.2