1.4 Reverse Process to Change Auto Sipper to 6 Cell Turret
Screw loosening (side view)

1.4.4 Detaching Front Panel

1.4.5 Attaching Sample Compartment Front Cover
Loosen the hexagonal screw on the coupling shaft using the
hexagonal wrench which is an accessory to the spectrophotometer.
Hexagonal
wrench
Coupling
shaft
Fig. 1-35 Detaching Coupling Shaft (2)
Remove the front panel used on the Auto Sipper.
Slide upward to dismount.
Fig. 1-36 Detaching Front Panel
In attaching the sample compartment front cover, check the tabs on
the front panel and on the instrument shown in Fig. 1-37. Engage the
left and right tabs on the sample compartment front cover
correspondingly with the tabs on the instrument, and then slide the
sample compartment front cover from the upper-side of the tab
downward (refer to Fig. 1-38).
1 - 20
Hexagonal
wrench
Screw loosening (top view)