Installation; Outline; Checking Before Mounting Auto Sipper; Installation Conditions - Hitachi UH5300 Instruction Manual

Table of Contents

Advertisement

1.1 Outline

1.2 Checking before Mounting Auto Sipper

1.2.1 Installation Conditions

1 INSTALLATION

The Auto Sipper is an instrument that is incorporated in the sample
compartment section of the Model UH5300 Spectrophotometer to
measure the absorbance or permeability of samples. Sample liquid is
sipped through the peristaltic pump into the flow cell in the Auto
Sipper. The sipping amount of sample can be adjusted according to
the time of rotation of the peristaltic pump.
Fig. 1-1 shows an external view of the Auto Sipper.
Fig. 1-1 External View of Auto Sipper
(1) Service temperature:
For measurement in the most stable conditions, it is
recommended to install the instrument in a room which is
air-conditioned at 20 to 25°C.
(2) Service humidity:
(Condensation is not allowed. Where the ambient temperature is
30°C or more, control the relative humidity to 70% RH or below.)
(3) Storage temperature: -20 to 60°C
(4) Storage humidity:
1 - 1
15 to 35°C
25 to 80% RH
5 to 95% RH
1.2.1