The buffer battery can be replaced if necessary.
WARNING
Risk of explosion
To reduce the risk of ignition of a flammable or explosive atmosphere, observe the
following:
► Do not charge or replace the battery underground or in explosion-hazard areas.
► Only use the battery charger specified by Dräger. The use of a different charger
nullifies the explosion protection certification of the gas detector.
● Insert the gas detector into the charger module.
If the gas detector is switched on, then it is automatically switched off after being
inserted.
6.9

Charge the battery with the multicharger

The following power supply units are available:
– 8321849 for max. 5 charging modules
– 8321850 for max. 20 charging modules
– 8321855 for max. 1 charging module when used with X-am 2x00/5x00 (vehicle
power supply unit)
NOTICE
Damage to charging modules
If several charging modules are disconnected at the same time, care must be taken
as otherwise the projecting tongue may break.
► Always connect or disconnect the charging modules individually and not in
groups.
|
Technical manual
X-am
®
2800
8321849
5x
2x
8321850
20x
10x
8321855
5x
2x
Maintenance
5x
20x
5x
27