Troubleshooting
5.1

Error

The sensor technology that can be used depends on the device type.
The first digit of the code represents the sensor channel:
Code
2xx
3xx
4xx
Code
102
103
104
105
108
109
110
111
112
113
115
122
123
124
125
132
141
148
149
16
Sensor channel
DrägerSensor CatEx SR
EC1
EC2
Cause
Customer's life time counter has
expired.
Gas detector defective.
Program code checksum error
Bump test interval expired
Gas detector defective.
Menu function cannot be per-
formed as an error has occurred.
Alarm element test faulty.
Faulty alarm element test:
Alarm LED
Faulty alarm element test:
Horn
Faulty alarm element test:
Vibration motor
Gas detector disabled by X-dock. Activate device with X-dock.
Internal device error
Hardware error
(reference voltage source)
Device temperature too high.
Device temperature too low.
Self-test failed
Vibration motor defective.
Ambient pressure too low.
DB version invalid for this firm-
ware.
Code
Sensor channel
5xx
EC3
6xx
EC4
Remedy
Reset life time counter with Dräger
CC-Vision.
Contact DrägerService.
Contact DrägerService.
Carry out bump test.
Contact DrägerService.
Determine fault code using the
Info menu and disable if neces-
sary.
Repeat the alarm element test
with X-dock.
Repeat the alarm element test
with X-dock.
Repeat the alarm element test
with X-dock.
Repeat the alarm element test
with X-dock.
Switch the gas detector off and on
again.
Contact DrägerService.
Contact DrägerService.
Operate gas detector within per-
mitted temperature range.
Operate gas detector within per-
mitted temperature range.
Contact DrägerService.
Contact DrägerService.
Operate gas detector within per-
mitted ambient pressure range.
Repeat the software update or
contact DrägerService.