Configuration

d. Carefully insert the new sensor.
e. Carefully plug the flexstrip into the bushing on the options board.
f. Close the sensor damper and make sure that it is positioned correctly.
7. Reattach the front and rear housing cradles and makes sure that the seals and
sensor dampers are in the correct positions.
8. Screw on the rear housing cradle with 4 screws (35 ±5 Ncm).
the two upper ones.
9. Follow the instructions in the sensor exchange wizard in the CC-Vision PC
software.
Next steps:
● After every sensor replacement, the fresh air calibration/zero calibration must
first be carried out, followed by the span calibration (mixed gas or single-gas).
● For the XXS O
test gas after the fresh air calibration.
6.12

Cleaning

The gas detector does not require any special care.
If it is heavily soiled, wash off the gas detector with cold water, using a sponge if
necessary. Dry the gas detector with a cloth.
NOTICE
Damage to the gas detector!
Harsh cleaning utensils (e.g. brushes), detergents and solvents can destroy the
dust and water filters.
► Only clean the gas detector with cold water and, where necessary, a sponge.
7
Configuration
After a change to the configuration (e.g. with the CC-Vision PC software), check
the basic functions of the gas detector (e.g. alarm elements).
32
Dräger recommends first tightening the two lower housing screws and then
, Dräger recommends testing the alarm feature with a suitable
2
For information on suitable cleaning agents and disinfectants and their
specifications, see document 9100081 at www.draeger.com/IFU.
|
Technical manual
X-am
®
2800