Warnings - Dräger X-am 2800 Technical Manual

Hide thumbs Also See for X-am 2800:
Table of Contents

Advertisement

Code
x01
x02
x03
x04
x05
x07
x08
x09
x12
x13
x14
x19
x21
x22
x23
x24
x36
x37
216
218
242
5.2

Warnings

The sensor technology that can be used depends on the device type.
The first digit of the code represents the sensor channel:
Code
2xx
|
Technical manual
X-am
Cause
No valid zero-point calibration.
No valid span calibration.
Measured value of sensor in neg-
ative range.
Sensor not connected or faulty.
Error during bump test.
Faulty rise time test.
Invalid value
Channel error
Calibration interval has expired.
Measurement invalid.
Sensor hardware error
If a channel error occurs on the
CO-channel when activating Tox-
icTwins, this error will be set on
the HCN channel.
Insufficient oxygen for the correct
functionality of the CatEx sensor.
Zero calibration failed (CatEx ther-
mal conduction range).
Span calibration failed (CatEx
thermal conduction range).
Gas detector incorrectly config-
ured by Dräger CC-Vision.
Device temperature too high.
Device temperature too low.
Thermal conduction calibration
interval for DrägerSensor CatEx
SR has expired.
Blocking alarm not plausible.
Sensor current for DrägerSensor
CatEx SR too low.
Sensor channel
DrägerSensor CatEx SR
®
2800
Remedy
Carry out a fresh air calibration /
zero-point calibration.
Carry out the span calibration.
Carry out a fresh air calibration /
zero-point calibration.
Connect sensor or check contacts.
Repeat the bump test.
Repeat the rise time test with X-
dock.
Contact DrägerService.
Carry out the span calibration.
Reinaugurate sensor.
Replace the sensor.
Recalibrate the CO-channel or
replace the sensor.
Operate the gas detector in atmo-
spheric conditions.
Carry out a fresh air calibration /
zero-point calibration.
Carry out the span calibration.
Replace sensor for affected chan-
nel with Dräger CC-Vision.
Operate gas detector within per-
mitted temperature range.
Operate gas detector within per-
mitted temperature range.
Perform span calibration for
DrägerSensor CatEx SR.
Calibrate sensor.
Contact DrägerService.
Code
Sensor channel
5xx
EC3
Troubleshooting
17