72 | DX-D 100 | Problem solving

Corner area of detector not exposed

Details
Cause
Solution
0187K EN 20191024 1609
A small area in the corner of the detector has not been ex-
posed.
When making an exposure with small source-image distance
(e.g. 1 m) and with the collimator in rotated position (e.g.
45°), the X-ray field does not reach to the outer corners of the
collimated area.
Increase the source-image distance.