44 | DX-D 100 | Introduction
For directions concerning X-ray radiation safety, electrical safety and
electromagnetic safety, refer to the DX-D 100 Mobile X-Ray Unit User Manual,
document 0188.
Topics:
•
Cleaning the system
•
Disinfection of the system
•
Emergency stop button
0187K EN 20191024 1609
being used and if the status of the DR Detector is ready for
exposure.
CAUTION:
The unit is equipped with a safety switch in the drive handle. If
the drive handle is released the unit will stop. In case of
unintended movement, do not try to correct the units'
movement via the drive handle but release the drive handle
immediately to stop the unit. If you experience unintended
movement, the unit has to be taken out of operation. Please
notify your Agfa service contact at once.
CAUTION:
Excessive ambient temperature may impact performance of DR
Detectors and cause permanent damage to the equipment. Refer
to the related user manual for environmental conditions for the
DR detector. If ambient temperature and humidity is outside the
specified range, do not operate the system or use air
conditioning. Warranty will be void if it is obvious that
operating conditions are not met.
CAUTION:
Damaged grid. Reduced image quality. Please handle the grids
with special care.