To Open the Analyzer Cover for Access to the Analyzer Sideplate
Analyzer cover
Upper RIS cover
Figure 20
Covers
Operation Manual
To Open the Analyzer Cover for Access to the Analyzer Sideplate
Open the analyzer cover
sideplate only. This is necessary during pump down or to access
the analyzer sideplate thumbscrews. You should not open the
analyzer cover for any other reason.
Procedure
1 Remove the upper RIS cover to expose the analyzer cover
latches. See
2 Open the latches on the analyzer cover at the front of the
instrument.
3 Swing the analyzer cover open.
Operating in Electron Impact (EI) Mode
(Figure
20) to access the analyzer
"To Remove the Upper RIS Cover"
3
on page 64.
65