4
Operating in Chemical Ionization (CI) Mode
To Switch from the EI Mode to CI Mode
84
Table 8
Default Tune Control Limits, used by CI autotune only
Reagent gas
Ion polarity
Abundance target
Peakwidth target
Maximum repeller
Maximum emission
current, µA
Max electron energy, eV
Notes for
Table
8:
• N/A Not available.
• Abundance target Adjust higher or lower to get desired
signal abundance. Higher signal abundance also gives higher
noise abundance.
• Peakwidth target Higher peakwidth values give better
sensitivity, lower values give better resolution.
• Maximum emission current Optimum emission current
maximum for NCI is very compound-specific and must be
selected empirically. Optimum emission current for
pesticides, for example, may be about 200 µA.
Methane
Positive
Negative
6
6
1x10
1x10
0.7
0.7
4
4
240
50
240
240
Ammonia
Positive
Negative
6
N/A
1x10
N/A
0.7
N/A
4
N/A
50
N/A
240
Operation Manual