Agilent Technologies 7200 Operation Manual page 124

Hide thumbs Also See for 7200:
Table of Contents

Advertisement

6
General Maintenance
To Clean the Ion Source
Ion focus lens
Figure 33
Extractor EI Source parts to be cleaned
Ion focus lens
Figure 34
CI Source parts to be cleaned
CAU TIO N
124
Extractor lens
Source body
The filaments, source heater assembly, and insulators cannot be
cleaned ultrasonically. Replace these components if major
contamination occurs.
3 If the contamination is serious, such as an oil backflow into
the analyzer, seriously consider replacing the contaminated
parts.
Source body
Repeller
Repeller