Restrictor and Column Installation

Restrictors and the column exit are connected to the splitter assembly using
internal nuts and SilTite ferrules. Install fittings on the column exit and one
end of each restrictor as described in

Install the column

1 Hang the analytical column on the column clips. The clips hold the outside
2 Connect the column to the inlet fitting.

Connect the splitter

1 Preswage the SilTite ferrules to the restrictors and columns. See
2 Connect the restrictors to the connectors on the splitter
3 Connect the restrictors to the detector inlet fittings. See your GC
4 Connect the column exit to the splitter. Tighten as you did the restrictors.
Arrange the tubing (restrictors and column) so that it does not touch the oven walls.
C A U T I O N
This could create a cold spot.
Installation and Operation Guide
of the wire basket that supports the column. Adjust the clips if necessary.
SilTite Ferrules"
on the CD for details.
Finger-tighten until just snug, then tighten with a wrench an additional 15°.
Install the back restrictor first.
documentation for details.
Splitter Configurations
"Swaging SilTite Ferrules"
3
on the CD.
"Swaging
(Figure
15).
35