APPENDIX 4: TROUBLESHOOTING GUIDE (continued)
PROBLEM
VITALERT 2000
running on battery
power.
Backup battery low.
Serial port not
communicating.
Strip chart recorder not
plotting.
38
POSSIBLE CAUSE
AC power cable disconnected.
Primary circuit breaker tripped.
The battery was in use recently;
the charge has been depleted.
Battery circuit breaker tripped.
Data cable disconnected.
VITALERT 2000 or device
improperly configured.
Printer out of paper.
No data available.
RETURN TO CD-ROM TABLE OF CONTENTS
REMEDY
Secure both ends of the AC
power cable.
Investigate cause and reset the
breaker.
Leave the VITALERT 2000
plugged into an active AC
receptacle for 16 hours.
Investigate cause and reset the
circuit breaker.
Secure all data cables.
Reconfigure port (see
Configuration section).
Insert paper into the strip chart
recorder.
Wait for valid data.