Imaging system overview

Detector (imaging unit)

The detector's (imaging unit's) function is listed below:
-
It converts the acquired X-ray image to a light signal using the fluorescent screen
and stores.
-
After reading the stored electric signal (image) from the sensor, it performs the A/D
conversion and temporarily stores it on the frame memory.
-
It reduces the electric signal during storing and transfers it to the CXDI PC through
the power box.

PCB-50Di board of the detector

The PCB-50Di board has the following functions:
-
The sensor driving control. Reading the digital image signals output from the AD
converter (PCB-50AD)
-
Communication, X-ray image transferring with the CXDI PC via the power box on
the Ethernet.
-
Synchronization of X-ray generation and image acquisition via the power box.
-
Control of the LED display.
-
Detection of the grid using the grid detecting sensor.
-
Relaying of the power supplied by the power box to the PCB-50AD.
-
Detection of an unusual temperature, error reporting and stopping of sensor driv-
ing.
-
Detection of the X-ray emission status.
-
Holding of the factory data set.