Backup and restore
Restore of a drive partition
Restore is required only if the imaging system is not working properly.
Restore c: + d: is performed when problems occur in the Win XP operating system or
imaging system application software.
After restoring a drive partition, the current organ programs and current customer adapta-
tions must be reinstalled using the last imaging system backup CD of the configuration
(refer to the "Backup/Restore of the Configuration" chapter).
Preliminary step:
-
The MOBILETT XP Digital system DVD is required.
-
The last MOBILETT XP Digital backup CD "Install#" is required.
Procedure
•
Switch off the MOBILETT XP Digital.
•
Open the top cover of the MOBILETT XP Digital, lay it sideways on a table or the like
and access the PC interior. Do not disconnect any cables.
•
Open the PC so that it is possible to use the PC keyboard.
•
Switch on the MOBILETT XP Digital and press "F2" on the keyboard during boot.
•
Enter the BIOS password "99999".
•
In the menu, verify that the first boot device is set to DVD ROM; change it if necessary.
Save changes and exit the menu.
•
Insert the CD MOBILETT XP Digital system DVD into the CD RW drive and boot from
the DVD.
•
Enter "2" and confirm with "Enter"; wait for a message to appear.
•
Following the message window,
•
Enter "E" to exit.
•
The configuration data must now be restored to reinstall all organ program and custom-
er adaptations. Refer to the "Backup/Restore of the Configuration" chapter.
•
Switch off the MOBILETT XP Digital.
Siemens AG
Medical Solutions
Access the BIOS mode of the computer.
The following menu opens:
- 1 - Back up the complete hard drive (c: + d:)
- 2 - Restore the complete hard drive (c: + d:)
- 3 - Back up the complete hard drive (c:
- 4 - Restore the complete hard drive (c:
- 5 - Use GHOST interactive
- E - Exit
Drive c: contains the operating system, drive d: contains the application soft-
ware.
A DOS window opens at the end of the restore operation.This concludes the
system restore of the image system.
SPR8-230.840.30.03.02
12.05
MOBILETT XP Digital
Page 59 of 68
CS SD 24
59