Calibration of the detector

4-
Calibration of the detector
3Calibration of the detector

General:

The detector is to be calibrated:
-
every 26 weeks.
-
In the event of an excessive amount of pixel dropouts on images displayed in the
image preview window.
During the calibration procedure 4 images must be exposed.

Prerequisite

Remove the grid if applicable
Center the tube unit above the detector.
Set an SID of 150cm
Use the most common X-ray parameter used by the customer, e.g. 70 kV,
1.8 mAs at the MOBILETT XP Digital.

Procedure

Select "SYSTEM/CALIBRATION" from the application UI.
Enter, e.g., the following parameters in the exposure data boxes = 4 exposure; 70KV;
1.8 mAs, SID 150cm
Be sure the detector is centered with respect to the tube without any objects between
them.
Click on "Start" to begin calibration.
Follow the instructions given by the system.
X
4 exposures are used to calibrate the system. No error messages should appear. After
the calibration procedure, the successful calibration is announced.
If the calibration error message "Too much/little dose" or "Collimator is used" is displayed,
perform calibration again after adjusting the X-ray exposure condition, for example by
increasing or reducing the mAs or KV settings.
If calibration fails again, try the self-test procedure described in this document or contact
the Uptime Support Center.
Exit the program via "Exit".
Restart the MOBILETT XP Digital
Siemens AG
Medical Solutions
Wait until the system is ready for calibration
SPR8-230.840.30.03.02