Error messages of the imaging system
Code
-507
-509
Siemens AG
Medical Solutions
Reply
Message
keys
RETRY
#1:Warning
CANCEL
#2:DICOM printer status error
#3:CXDI has aborted DICOM transmission because of a FAIL-
URE in the DICOM printer status. Check receiving device.
[Cause]
DICOM printer status was received properly. However, the tar-
get reported a FAILURE as the printer status, and image trans-
fer was consequently cancelled.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be transferred again.
When CANCEL is selected, the transfer will be cancelled after a
confirmation message is displayed. Re-transfer is possible by
sending the image from the study list.
RETRY
#1:Warning
CANCEL
#2:External storage error:
#3:Selected directory or disk for storage can only be read.
[Cause]
The external storage medium is read-only.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be stored. When CAN-
CEL is selected, the storage of the image to an external storage
medium will be cancelled after a confirmation message is dis-
played. Re-transfer is possible by sending the image from the
study list again.
SPR8-230.840.30.03.02
12.05
MOBILETT XP Digital
Page 43 of 68
CS SD 24
43