Error messages of the imaging system
Error Code
-507
-509
-510
-512
-514
-515
-516
-517
-518
-519
Siemens AG
Medical Solutions
Error Message
DICOM Printer Status Error: FD
has aborted DICOM transmis-
sion because of a FAILURE in
DICOM printer status. Check
receiving device.
External Storage Error: Selected
directory or disk for storage is
read-only.
External Storage Error: Selected
directory or disk for storage does
not exist. Or image cannot be
saved on external storage,
because the external storage or
directory is not properly set.
External Storage Error: Cannot
recognize the external storage.
Check disk settings and retry.
External Storage Error: Exposed
images cannot be stored
because capacity of folder parti-
tion is smaller than size of
exposed images. Use a new
disk, erase folder contents, or
increase folder size and retry.
Internal Storage Error: Selected
directory for storage is
read-only.
Internal Storage Error: Selected
directory for storage does not
exist, or attribution is incorrect.
Internal Storage Error: Pro-
cessed image cannot be stored
because capacity of internal
storage memory is insufficient.
Internal Storage Error: DICOM
image cannot be stored because
capacity of internal storage
memory is insufficient.
External Storage Error: Capac-
ity of external storage memory is
insufficient. Retry after com-
pressing data.
SPR8-230.840.30.03.02
12.05
Remedy
•
After eliminating the printer error,
touch RETRY.
•
Or touch Cancel and cancel the
transfer.
•
Replace the disk with a writable disk
and select RETRY.
•
Or select Cancel. Or contact the
network administrator.
•
Touch OK and shut down the sys-
tem.
•
If these errors occur, contact a Sie-
mens representative or distributor.
•
Touch RETRY after the green mark
has disappeared from the message
area of the exposure screen.
•
Select RETRY to save the image to
the disk again. Or select Cancel.
•
Replace the disk with a writable disk
and select RETRY.
•
Or select Cancel. Or contact the
network administrator.
Page 49 of 68