Hardcopy Messages; Error Return Values For Dicom Print Services - Siemens MOBILETT XP Digital Troubleshooting Manual

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

Error messages of the imaging system

Hardcopy messages

Error return values for DICOM print services

Tab. 11
DICOM Print Service messages
Sample error
Success
Invalid parameter
The server has not
started up.
Server fails to
start.
Siemens AG
Medical Solutions
Return value
0x00
0x00000001
0x00180012
SPR8-230.840.30.03.02
12.05
CXDI description/ log output
There is no log output in this case.
CXDI description:DICOM connection error
(CCRTRANS_ERR_DICOMPARAM)
------------------------------------------------------------------
The log output in this case is shown below.
- -s Silent mode; do not print results of all print
commands
- -t) Set called AE title to title in Association RQ
- -v Use verbose mode for DUL and SRV facili-
ties
- -x Canon Hidden Special Mode
- node The host name that is running a print
server
- port TCP/IP port number of print server
- file One or more files that contain preformat-
ted images for printing
- image) A list of one or more images to send
CXDI description: DICOM connection error
(CCRTRANS_ERR_NOT_CONNECT
The log output in this case is shown below.
-------------------------------------------------------------------
-18-135933[d2]ERR: d0012 Attempt to connect to
unknown host: test
18-135933[d2]ERR: 130012 Peer aborted Associ-
ation(or never connected)
18-135933[d2]ERR:180012 Failed to establish
association
18-132600[127]ERR:60012 TCP Initialization
error: Invalid argument
18-132600[127]ERR:130012 Peer aborted Associ-
ation (or never connected)
18-132600[127]ERR:180012 Failed to establish
association
Page 53 of 68