Siemens MOBILETT XP Digital Troubleshooting Manual page 20

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

20
Code
-116
-5002
MOBILETT XP Digital
Error messages of the imaging system
Reply
Message
keys
OK
#1:Alert
#2:Sensor unit:Self-check error
#3:Detected an error.
#4:ER(n) (xxxx) n: Sensor index (1 to 4) xxxx: Error code
[Cause]
An undefined error has been found in the sensor A/D board.
[Operation after the reply key is pressed] Nothing
OK
#1:Alert
#2:OPU-CCR RPC Error
#3:OPU-CCR RPC Error
[Cause]
For some reason, the CCR software has shut down.
[Operation after the reply key is pressed]