Error messages of the imaging system
Error Code
-5102
-5202
-5203
-5206
Siemens AG
Medical Solutions
Error Message
Grid Error: Remove grid for this
exposure.
Grid ID Error: Replace with the
recommended grid (name of
grid). Installed grid: (name of
grid).
Grid ID Error: This is a non-grid
mode. Remove the grid.
Grid ID Error: This is a fixed-grid
mode. Attach the grid.
SPR8-230.840.30.03.02
12.05
Remedy
•
Touch OK and remove the grid from
the detector unit.
•
Touch OK and check the grid in the
detector unit.
•
Touch OK and remove the grid from
the detector unit.
•
Touch OK and install the grid on the
detector unit.
Page 39 of 68
CS SD 24
39
MOBILETT XP Digital