Error messages of the imaging system
Error Code
-5102
-5202
-5203
-5206
Siemens AG
Medical Solutions
Error Message
Grid Error: Remove grid for this
exposure.
Grid ID Error: Replace with the
recommended grid (name of
grid). Installed grid: (name of
grid).
Grid ID Error: This is a non-grid
mode. Remove the grid.
Grid ID Error: This is a fixed-grid
mode. Attach the grid.
SPR8-230.840.30.03.02
12.05
Remedy
Touch OK and remove the grid from
the detector unit.
Touch OK and check the grid in the
detector unit.
Touch OK and remove the grid from
the detector unit.
Touch OK and install the grid on the
detector unit.