Use the initialization code when the normal code in unable to start for some reason or
when you have lost the Ethernet settings. Using the initialization code sets the Ether-
net-related settings to the default settings and allows you to perform the start-up operation
for the initialization code.
In this case, the connection can be made by setting the Ethernet settings in the control PC
to the 50G default (*2).
This default connection allows you to upgrade the firmware again and also to check and
set the Ethernet settings again in order to recover the system.
Fig. 6:
Initialization switch: Switch 1 mounted on PCB-50 XRY
Default settings: IP Address=192.168.100.11 (factory setting)
MOBILETT XP Digital
Power box installation switch
SPR8-230.840.30.03.02
12.05