Error messages of the imaging system

Attention

Tab. 9
Service Attention Table
Code
-506
-508
-511
Siemens AG
Medical Solutions
Reply
Message
keys
OK
#1:Attention
#2:DICOM response warning
#3:Although transfer finished properly, the receiving device
reported a warning in the DICOM response. Check receiving
device.
[Cause]
DICOM response was received properly. However, the target
reported a warning as a response.
[Operation after the reply key is pressed]
Checks only if the system is OK because the system is normal.
OK
#1:Attention
#2:DICOM printer status warning
#3:Although transfer finished normally, the receiving device
reported a WARNING in the DICOM printer status. Check
receiving device.
[Cause]
DICOM printer status was received properly. However, the tar-
get reported a WARNING as the printer status.
[Operation after the reply key is pressed]
Only checks if the system is OK.
OK
SPR8-230.840.30.03.02
12.05
MOBILETT XP Digital
Page 51 of 68