42
Code
-503
-504
-505
MOBILETT XP Digital
Error messages of the imaging system
Reply
Message
keys
RETRY
#1:Warning
CANCEL
#2:DICOM transfer error
#3:Error has occurred during association.
[Cause]
Network was cut off during communication.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be transferred again.
When CANCEL is selected, transfer will be cancelled after a
confirmation message is displayed. Re-transfer is possible by
transferring the image from the study list.
RETRY
#1:Warning
CANCEL
#2:DICOM parameter error
#3:DICOM transfer parameter is wrong.
[Cause]
DICOM transfer parameters are not correct.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be transferred again.
When CANCEL is selected, transfer will be cancelled after a
confirmation message is displayed. Enter the correct parame-
ters and transfer the image from the study list again.
RETRY
#1:Warning
CANCEL
#2:DICOM response error
#3:Receiving device reported an error in DICOM response.
Check receiving device.
[Cause] DICOM response was received properly. However, the
target reported an error as a response.
[Operation after the reply key is pressed] When RETRY is
selected, the image will be transferred again.When CANCEL is
selected, transfer will be canceled after a confirmation message
is displayed. Re-transfer is possible by sending an image from
the study list.
SPR8-230.840.30.03.02
12.05
Page 42 of 68