Siemens MOBILETT XP Digital Troubleshooting Manual page 28

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

28
Code
-305
-306
-310
MOBILETT XP Digital
Error messages of the imaging system
Reply
Message
keys
OK
#1:Error
#2:Calibration error
#3:Collimator may be used.
[Cause]
This error is not currently used.
[Operation after the reply key is pressed]
Perform imaging again.
OK
#1:Error
#2:QA process error
#3:Image analysis error. Adjust image process parameter man-
ually.
[Cause]
Feature extraction may have failed.
[Operation after the reply key is pressed]
The user must process the image manually.
OK
#1:Error
#2:REX value warning
#3:REX is 999.
It is under the minimum limitation value 9999. Please reconfirm
the generator setting or exposure condition. If the condition is
correct, adjust the image processing.
[Cause]
This error occurs when the REX value acquired by image pro-
cessing after the imaging is less than the limit value.
[Operation after the reply key is pressed]
Nothing
[Note]
999: REX value obtained by image processing 9999: Lower
limit value
SPR8-230.840.30.03.02