26
Code
-129
-143
-144
MOBILETT XP Digital
Error messages of the imaging system
Reply
Message
keys
OK,
#1:Error
0xEFF00
#2:Sensor unit: Data transmit error
xEFEF
#3:Failed to transmit the data.
[Cause]
Internal DMA has exceeded the time limit while reading the
reduced image. Full image data transferring has been inter-
rupted.
[Operation after the reply key is pressed]
No action required. (Usual operation)
0xEFF00xEFEF
n.a.
OK
#1:Error
#2:Calibration table error
#3:Perform calibration.
[Cause]
No gain calibration table has been prepared. Or the serial num-
ber of the connected sensor unit is different.
[Operation after the reply key is pressed]
Perform calibration to make a gain calibration table.
OK
#1:Error
#2:Sensor unit configuration
#3:Incorrect senor unit configuration. Call a service person.
#4:Invalid serial no.(n)(eff5 or eff6)n: Sensor index (1 to 4)
[Cause]
0xeff5: Caused by the difference in serial number and DP num-
ber. 0xeff6: Caused by a false serial number being written.
[Operation after the reply key is pressed]
When the OK button is pressed, the sensor is disconnected. A
service person needs to be called in. (Maintain unusual state,
and do not access.)
SPR8-230.840.30.03.02
12.05
Page 26 of 68