Analog; Display; Response Time - E+E Elektronik EE75 User Manual

Air / gas velocity sensor
Hide thumbs Also See for EE75:
Table of Contents

Advertisement

4.2
The "Analog" tab allows free configuration and
scaling of the twoanalogue outputs
Range:
The drop-down input field is used either to select a
standardised output signal (0-5 V, 0-10 V, 0-20 mA,
4-20 mA) or to specify a user-defined current/
voltage output range (e.g. 1-9 V).
Measurement
Specifies the physical quantities the outputs will
value:
represent.
If the measurement value "volumetric flow rate" is
selected, the cross section has to be defined (see
chapter 4.5 Cross section).
Upper / Lower
Specifies the required display range. The limits must
Limit:
lie within the maximum span as indicated.
Physical quantity:
Allows a user to choose whether measuring results are displayed and output in
SI units (m/s; °C; m³/min) or US units (ft/min.; °F; ft³/min.).
4.3
With an optional display, the following can be set on the Display tab:
"Display Mode" drop-down input field:
- Single-line display
- Two-line display (factory setting)
"Backlight on" check box
- Activated = ON
- Not activated = OFF
4.4
As described under Hardware, chapter 5.2 - Jumpers for Setting the Velocity Response Time, three
different response times can be set.
In the factory settings, the following response times are assigned to the 3 jumper positions:
Jumper at position 1: τ
Jumper at position 2: τ
Jumper at position 3: τ
Users can also assign a response time of between 1.5 and 40 s to each jumper position using the
configuration software.
Example:
Jumper at position 1: changes from 2 s (factory setting) to 35 s.
The green pilot lamp next to the jumper position indicates which jumper position is
currently set and consequently active.
Note: If the response time of the jumper currently active is changed anduploaded using the
"Write" function, the sensor's velocity response is switched without changing the hardware
(jumper position).
Configurations Software

Analog

Display

Response Time

= 2 s
90
= 4 s
90
= 10 s
90
17

Advertisement

Table of Contents
loading