Duct sensor for co2, humidity and temperature (13 pages)
Summary of Contents for E+E Elektronik EE300Ex
Page 1
User Manual EE300Ex HUMIDITY / TEMPERATURE SENSOR BA_EE300Ex_e // V1.11.2 // Modification rights reserved // 194916...
Page 2
E+E Elektronik Ges.m.b.H. doesn‘t accept warranty and liability claims neither upon this publication nor in case of improper treatment of the described products. The document may contain technical inaccuracies and typographical errors. The content will be revised on a regular basis. These changes will be implemented in later versions. The described products can be improved and changed at any time without prior notice.
This symbol indicates safety information. It is essential that all safety information is strictly observed. Failure to comply with this information can lead to personal injuries or damage to property. E+E Elektronik® assumes no liability if this happens. This symbol indicates instructions.
Environmental Aspects Products from E+E Elektronik® are developed and manufactured observing of all relevant requirements with respect to environment protection. Please observe local regulations for the device disposal. For disposal, the individual components of the device must be separated according to local recycling regulations.
• Mixing ratio • Water vapour partial pressure e Besides measurement in the air, the EE300Ex-M1 with ATEX or IECEx approval is also suitable for measuring water content (X) in ppm and water activity (aw) in isolation, lubrication and hydraulic oils.
Ex Certificate. The “hazardous label” shows the Ex marking and the certificate number. See the examples below. EE300Ex with IECEx, USA, Canada, Korea or Japan labeling must not be installed in the European Union.
2.3 Certification EUROPE: The EE300Ex sensor fulfills the ATEX Directives on intrinsically safe operating equipment. Applied standards for ATEX: • EN 1127-1:2011 • EN 60079-0:2012 • EN 60079-11:2012 The EU-Type Examination has been carried out by TÜV SÜD Product Service GmbH.
Page 9
• T3 temperature class based on -70°C (-94°F) ≤ Ta ≤ 175°C (347°F) • T2 temperature class based on -70°C (-94°F) ≤ Ta ≤ 200°C (392°F) • T1 temperature class based on -70°C (-94°F) ≤ Ta ≤ 200°C (392°F) User Manual EE300Ex Humidity / Temperature Sensor...
Page 10
Sensor with display Class I, Division 1, Groups CD; T4 Ta = -40°C to +60°C; Entity – M1_139080 Class I, Division 2, Groups ABCD; T4 Ta = -40°C to +60°C; Entity – M1_139080 User Manual EE300Ex Humidity / Temperature Sensor...
Page 11
• T1 temperature class based on -70°C (-94°F) ≤ Ta ≤ 200°C (392°F) KOREA: Applied law for South Korea: Article 34 of the Occupational Safety & Health Act The Certificate of Conformity has been carried out by KC. User Manual EE300Ex Humidity / Temperature Sensor...
Page 12
For use EE300Ex with remote probe in Zone 0 the remote probe must be installed in a metallic grounded protective tube. – EE300Ex with the connector option may not be used in dust hazard area and in gas hazard area Zone 0 for group IIC.
Entity parameters: Ui = 28Vdc; Ii = 100mA; Pi = 700mW; Ci = 2.2nF; Li = 0mH Ex-Designation Ex ia IIC T4 Gb (Ta = -40°C to 60°C) No installation in Ga permitted. EE300Ex must be mounted in Zone 1 or Zone 2. No dust certification or temperature probe without humidity. No plastic filter caps allowed.
Requirements for the installation of overvoltage protection devices are set in European Normative EN 60079-25. For the U.S., Canada refer to ANSI/ISA RP12.6.01, NEC and CEC. The manufacturing date of each EE300Ex is shown on the hazardous product label, at the bottom right corner, as follows: WWYYYY WW ..week of the year...
3.3 Assembly in Category 1 (Zone 0 / 20); Division 1 Only intrinsically safe power supply devices are approved to supply EE300Ex in category 1 or Division 1. In areas belonging to gas group IIC or Class I, Division 1, Group A, B, it must be ensured that during installation and operation, the possibility of impact and friction sparks has been excluded in rarely occurring fault situations.
Page 16
Class II, III and Class I, Division 1, Group A, B Control intrinsically safe sensor supply unit 4-20mA ia Power supply EE300Ex (with remote probe) 2 channels via intrinsically safe power supply device: User Manual EE300Ex Humidity / Temperature Sensor...
Assembly in Categories 2 and 3 (Zone 1 , 2 / 21 , 22); Division 2 Only intrinsically safe power supply devices and protective barriers are approved to supply EE300Ex in category 2 and 3 or Division 2.
EE300Ex (with remote probe) 2 channels via intrinsically safe power supply device: Hazardous location T4...T1 Unclassified location Zone 0 / 20 Zone 1 2 / 21 22 Division 1 Division 2 Class I,II,III Class I,II,III Group A, B, C, D, E, F, G, H...
• Mark the nut (A) at 6 o‘ clock position. • Hold the screw connection body (B) tight and tighten the nut (A) with 1 ¼ turns to 9‚ o clock position. User Manual EE300Ex Humidity / Temperature Sensor...
An excessive tightening torque may cause permanent deformation of the clamping sleeve and of the probe. This makes the removal and re-mounting difficult or even impossible. User Manual EE300Ex Humidity / Temperature Sensor...
3.5.4 The Optional Probe Retraction Tool The operating instructions of the probe retraction tool must be strictly observed. The probe of the EE300EX model T22 can be installed into a pressurized environment up to 250 bar using the sensor retraction tools ZM-WA-025-040-EST or BG-WA-103-045-EST.
Installation in Canada shall comply with the Canadian Electrical Code (CEC). The EE300Ex sensor is a loop powered (2-wire) device, both channels (outputs) are galvanic isolated. For proper operation, Channel 1 (CH1) must always be connected. Channel 2 (CH2) may be connected only if necessary.
4.3 Grounding and Potential Equalization The EE300Ex shall be integrated into the potential equalization to avoid hazards from electrostatic charges. The grounding shall comply with EN60079-14, EN60079-25 or IEC60079-14, IEC60079-25. A remote probe shall also be grounded using a screw connection with a maximum of 1 MΩ in the potential equalization.
4.4 Connection Cable A shielded cable shall be used. The shield must be grounded only at one end, at the EE300Ex side. The connection cable shall comply to the following specifications for ATEX, as required by EN60079-14 and EN60079-25: •...
EN 60079-25:2010 13.1. CH1 and CH2 are galvanically isolated. The proof of intrinsic safety shall be performed with appropriate equipment. Example: Installation in Zone 0 of the EE300Ex without display Connection cable length: 300 m (984 ft) Explosion group: IIC Temperature class: T4 CH1 and CH2 connected by one single cable.
4.7 Configuration Adapter The on-board service interface is dedicated for the EE300EX configuration and for the RH and T adjust- ment. This is possible by the optional EE-PCA Product Configuration Adapter and HA011068 connec- tion cable. Refer to datasheet EE-PCA and HA011068 user manual.
The configuration or adjustment of EE300Ex may not be performed in the hazard area. CH1 and CH2 must be disconnected while using the configuration adapter. When the front section has been removed from the hazard area, e.g. for calibration, the empty bottom section shall be protected against dirt and electrostatic charge with the blind front cover HA011401 (see accessories).
Error 1 = RH sensing element damaged Error 2 = Condensation on the RH sensor element Error 3 = T sensing element damaged Error 4 = Short circuit at the T sensing element User Manual EE300Ex Humidity / Temperature Sensor...
1) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement). User Manual EE300Ex Humidity / Temperature Sensor...
1) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement). User Manual EE300Ex Humidity / Temperature Sensor...
10 IECEX Certification of Conformity - COC for more information see http://www.iecex.com/ or our website http://www.epluse.com/ee300ex User Manual EE300Ex Humidity / Temperature Sensor...
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: information@fmapprovals.com www.fmapprovals.com F 347 (Mar 16) Page 1 of 7 User Manual EE300Ex Humidity / Temperature Sensor...
Page 38
Zone 20, AEx ia IIIC T80°C Da; Entity – M1_1309080; IP65 12. Description of Equipment: General - The EE300Ex transmitter is designed for gauge measurements of temperature and humidity in air. All signal outputs are available on two 4 to 20 mA analog outputs.
Page 39
US Certificate Of Conformity No: FM17US0302X Ratings - The EE300Ex transmitter operates at 6.4-28 Vdc (700mW). The transmitters are rated for use in an ambient temperature range of -40°C to +60°C. The transmitter probe is rated for use in a process temperature range of -70°C to +200°C.
Page 40
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: information@fmapprovals.com www.fmapprovals.com F 347 (Mar 16) Page 4 of 7 User Manual EE300Ex Humidity / Temperature Sensor...
Page 41
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: information@fmapprovals.com www.fmapprovals.com F 347 (Mar 16) Page 5 of 7 User Manual EE300Ex Humidity / Temperature Sensor...
Page 42
Zone 0. 2. The EE300Ex Remote Probe is approved for in air applications only. 3. The serial software configuration port of the EE300EX may only be used with the Product Configuration Adapter (PCA) and Connection cable (HA011068). Equipment Group II: EE300Ex with display 1.
Page 43
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: information@fmapprovals.com www.fmapprovals.com F 347 (Mar 16) Page 7 of 7 User Manual EE300Ex Humidity / Temperature Sensor...
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: information@fmapprovals.com www.fmapprovals.com F 348 (Mar 16) Page 1 of 7 User Manual EE300Ex Humidity / Temperature Sensor...
Page 45
Zone 20, Ex ia IIIC T80°C Da; Entity – M1_1309080; IP65 12. Description of Equipment: General - The EE300Ex transmitter is designed for gauge measurements of temperature and humidity in air. All signal outputs are available on two 4 to 20 mA analog outputs.
Page 46
Canadian Certificate Of Conformity No: FM17CA0154X Ratings - The EE300Ex transmitter operates at 6.4-28 Vdc (700mW). The transmitters are rated for use in an ambient temperature range of -40°C to +60°C. The transmitter probe is rated for use in a process temperature range of -70°C to +200°C.
Page 47
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: information@fmapprovals.com www.fmapprovals.com F 348 (Mar 16) Page 4 of 7 User Manual EE300Ex Humidity / Temperature Sensor...
Page 48
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: information@fmapprovals.com www.fmapprovals.com F 348 (Mar 16) Page 5 of 7 User Manual EE300Ex Humidity / Temperature Sensor...
Page 49
Zone 0. 2. The EE300Ex Remote Probe is approved for in air applications only. 3. The serial software configuration port of the EE300EX may only be used with the Product Configuration Adapter (PCA) and Connection cable (HA011068). Equipment Group II: EE300Ex with display 1.
Page 50
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail: information@fmapprovals.com www.fmapprovals.com F 348 (Mar 16) Page 7 of 7 User Manual EE300Ex Humidity / Temperature Sensor...
14 Certificates Korea and Japan For more information please contact the E+E sales support. User Manual EE300Ex Humidity / Temperature Sensor...
Page 55
User Manual EE300Ex Humidity / Temperature Sensor...
Page 56
HEADQUARTERS SUBSIDIARIES E+E Elektronik Ges.m.b.H. E+E Elektronik China E+E Elektronik Germany E+E Elektronik Korea Langwiesen 7 18F, Kaidi Financial Building, Schöne Aussicht 8 C Suite 2001, Heungdeok IT 4209 Engerwitzdorf No.1088 XiangYin Road 61348 Bad Homburg Valley Towerdong, 13, Austria 200433 Shanghai Tel.:...