Agilent Technologies 1260 Infinity User Manual page 157

Standard autosampler
Hide thumbs Also See for 1260 Infinity:
Table of Contents

Advertisement

11
Hardware Information
Firmware Description
Firmware Updates
Firmware updates can be done using your user interface:
• PC and Firmware Update Tool with local files on the hard disk
• Instant Pilot (G4208A) with files from a USB Flash Disk
• Agilent Lab Advisor software B.01.03 and above
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
PPPP is the product number, for example, 1315AB for the G1315A/B DAD,
R the firmware revision, for example, A for G1315B or B for the G1315C DAD,
VVV is the revision number, for example 102 is revision 1.02,
XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware in
chapter "Maintenance" or use the documentation provided with the
Firmware Update Tools.
Update of main system can be done in the resident system only. Update of the resident
N O T E
system can be done in the main system only.
Main and resident firmware must be from the same set.
Figure 20
Firmware Update Mechanism
Agilent 1260 Infinity Standard Autosampler User Manual
157