Agilent Technologies 1260 Infinity User Manual page 115

Standard autosampler
Hide thumbs Also See for 1260 Infinity:
Table of Contents

Advertisement

3
Insert the new needle-seat assembly. Press the seat firmly
into position.
5
Use Down to position the needle approximately 2 mm
above the seat.
Next Steps:
7
On completion of this procedure: Install the front cover.
8
Select End in the maintenance function Change Needle (see
Agilent 1260 Infinity Standard Autosampler User Manual
Exchanging the Needle-Seat Assembly
4
Connect the seat-capillary fitting to port 5 of the injection
valve.
6
Ensure the needle is aligned with the seat. If required,
bend the needle slightly until the needle is aligned
correctly.
"Change Needle"
on page 60).
8