Transport And Set Up; Location - Thermo Scientific Sorvall BP 8 Instruction Manual

Hide thumbs Also See for Sorvall BP 8:
Table of Contents

Advertisement

Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges
THE ACTIVITIES DESCRIBED IN THIS CHAPTER ARE FOR SERVICE TECHNICIANS.

Transport and Set Up

CA UT I O N
Make sure that the centrifuge and the packaging is not damaged. Speak to the shipping company and
Thermo Fisher Scientific immediately if damage is found.

1. Location

WA R N I N G
Risk of impact by moving centrifuge.
The centrifuge can crush into objects and persons in a radius of 30 cm when spinning.
Keep a safety zone of 30 cm around the centrifuge for safe operation.
Make sure that no one is in the safety zone while the centrifuge spins.
CA UT I O N
UV rays decrease the stability of plastics.
Do not expose the centrifuge, rotors and plastic components to direct sunlight.
Operate the centrifuge only inside a room.
The set-up location must fulfill these requirements:
Keep a safety zone of minimum 30 cm
ƒ
Persons and hazardous substances must be kept out of this safety zone while centrifuging.
ƒ
The supporting structure must:
»
be stable and free of resonance,
»
be applicable for horizontal setup of the centrifuge,
»
hold the weight of the centrifuge.
Do not expose the centrifuge to heat and strong sunlight.
ƒ
The set-up location must be well-ventilated at all times.
ƒ
THE ACTIVITIES DESCRIBED IN THIS CHAPTER ARE FOR SERVICE TECHNICIANS.
Transport and Set Up
(Figure I–1 auf Seite
I–2, green area) around the centrifuge.
I–1
I