II. Transport and Set Up

1. Before Setting Up
1�
Check the centrifuge and the packaging for any shipping damage� Inform the shipping
company and Thermo Fisher Scientific immediately if any damage is discovered�
2�
Remove the packaging�
NOTICE Dispose the packaging� Do not reuse it�
3�
Check, if the items supplied are complete
If the items supplied are incomplete, please contact Thermo Fisher Scientific�
2. Location
CAUT ION
The centrifuge is only to be operated indoors�
The set-up location must fulfill the following requirements:
ƒ
A safety zone of at least 30 cm must be maintained around the centrifuge� People and
hazardous substances must be kept out of this safety zone while centrifuging�
Thermo Scientific Sorvall ST 8FR Centrifuge
("Items Supplied" on page
UV rays reduce the stability of plastics.
Do not subject the centrifuge, rotor and plastic accessories to direct
sunlight.
16
6)�