Thermo Scientific Sorvall BP 8 / 16 and Sorvall BIOS 16 Centrifuges and Heavy Duty Centrifuges
3. Troubleshooting by Guide
CAU T I ON
If problems occur other than those listed in this table, service technician must be contacted.
If an error occurs, follow the instructions on the touchscreen. Error messages have a red colored background.
Figure V–3: Example for an error message
Error Message
E-1 to E-97
E-98
E-99
Table V–1: Error messages
Troubleshooting
Description
Centrifuge can not be operated.
Centrifuge run can not be started or
centrifuge is braking.
V–3
Troubleshooting
Switch the centrifuge off and on again.
Make sure that the correct windshield lid
is inserted into the holder.
Move the drive shaft slightly towards the
front of the centrifuge.
If the error message still shows, contact
a service technician.
Imbalance detected.
Check the loading of the rotor.
Check the lubrication of the trunnion
bolts of the rotor body.
Switch the centrifuge off and on again.
If the error message still shows, contact
a service technician.
Switch the centrifuge off and on again.
If the error message still shows, contact
a service technician.