5 Reuse and cleaning

The product is reusable. Before using the
product again for a new user, it is
recommended to clean the product. Follow the
instructions below for cleaning and
reconditioning. Consult your hygiene manager
or Care of Sweden for help and instructions if
you are unsure.
CONTROL UNIT
Wipe off the control unit with a damp
cloth and mild cleaning solution for
this purpose: such as alcohol with or
without tenside surfactants, oxidizing solutions,
chlorine max 1 % or hydrogen peroxide max 1,5
%.
MATTRESS COVER
Covers can be wiped off with a mild cleaning
solution for this purpose such as: alcohol with
or without tenside surfactants, Isopropanol 70
%, oxidizing solutions, chlorine max 1 % (max
10% for cover Stone) or hydrogen peroxide max
1,5 %.
Excrement and blood stains must be removed
as soon as possible using cold water. Carefully
follow local instructions and the instructions for
the detergent.
Multi‐layer covers should be divided before
washing. The parts should be washed with
similar colours.
Chlorine and phenol‐based cleaners
could adversely affect the PU surface
and should be avoided. If chlorine is used, we
recommend a mixture of max 1 % (max 10% for
cover Stone).
INNER COVER
Clean the affected area with a mild cleaning
solution for this purpose such as: alcohol with
or without tenside surfactants, Isopropanol 70
%, oxidizing solutions, chlorine max 1 % or
hydrogen peroxide max 1,5 %.
MATTRESS
1. Disconnect the tube connector from
the control unit and remove the air
from the mattress.
2. Remove the covers.
3. Move the mattress to a clean area that is
suitable for cleaning.
4. Wipe off the air cells, tubes, and the CPR
module with a mild cleaning solution for this
purpose such as: alcohol with or without
tenside surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % or hydrogen
peroxide max 1,5 %. Let it dry.
5. Wipe off the work surfaces using a suitable
detergent and/ or disinfectant.
6. Put the mattress together. If air cells for any
reason have become loose from the tubes,
these must be replaced according to the
drawing in section 4.2.
Note!
• Check the hygiene cover, air cells and hoses
each time the product is cleaned. If damaged,
it must be replaced or repaired.
• Also check the control unit, tube connectors
and power cable during cleaning. Damaged
parts must be replaced or repaired.
• The air cells can be divided using a quick-
release connector, making it easier to clean
the air cells.