4. Reuse, cleaning and reconditioning

The product is reusable. When reusing, it is important to follow the instructions below for
cleaning and reconditioning. Before using the product again for a new user, it is recommended
to clean the mattress in accordance with the instructions below. Consult your hygiene
manager or Care of Sweden for help and instructions if you are unsure.
Cleaning and reconditioning should otherwise be performed as required.
NOTE!
Check the hygiene cover each time the product is cleaned. If damaged, it must be
replaced or repaired.
4.1 Covers
MATTRESS COVER
Covers can be wiped off with a mild cleaning solution for this purpose such as: alcohol
with or without tensides, Isopropanol 70 %, oxidating solutions or chlorine max 1 %.
Excrements and blood spots must be removed immediately with cold water. Carefully
follow local instructions and usage instructions for the cleaner.
Multi-layer covers should be divided before washing.
Chlorine and phenol-based cleaners could have a negative impact on the PU surface
and should be avoided. If chlorine is used, we recommend a mixture of max. 1 %.
INNER COVER
Clean the affected area with a mild cleaning solution for this purpose such as: alcohol with or
without tensides, Isopropanol 70 %, oxidating solutions or chlorine max 1 %.
4.2 Foam core
Clean the affected area with a mild cleaning solution for this purpose such as: alcohol
with or without tensides, Isopropanol 70%, oxidating solutions or chlorine max 1%.
Press out the water carefully. Let the foam core dry in a warm, ventilated area, without direct
exposure of sunlight. The foam core needs to be completely dry before it is used again.
4.3 Air cells
1. Remove the covers.
2. Carefully remove the air cells from the foam core. Remove the cell holder from the cells.
3. Place the air cells and the tubes on a clean area which is suitable for washing.
4. Wipe off the cells, all tubes with a mild cleaning solution for this purpose such as: alcohol
withor without tensides, Isopropanol 70 %, oxidating solutions or chlorine max 1 %.
See instructions in chapter 4.1 for cleaning covers.
Wash with similar colours.
10