7

Reuse and cleaning

The product is reusable. Follow the instructions
below for cleaning and reconditioning. It is
recommended to clean the mattress between
users. Consult your hygiene manager or Care of
Sweden for help and instructions if you are
unsure. Cleaning and reconditioning should
otherwise be performed as required.

Cleaning

CONTROL UNIT
Wipe off the control unit with a damp
cloth and mild cleaning solution for this
purpose: such as alcohol with or without
tenside surfactants, oxidizing solutions, chlorine
max 1 % or hydrogen peroxide max 1,5 %.
MATTRESS COVER
Covers can be wiped off with a mild
cleaning solution for this purpose such
as: alcohol with or without tenside
surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % (max 10% for cover
Stone) or hydrogen peroxide max 1,5 %.
Excrement and blood stains must be removed as
soon as possible using cold water. Carefully
follow local instructions and the instructions for
the detergent.
Multi-layer covers should be divided before
washing. The parts should be washed with
similar colours.
For chlorine, we recommend a mixture
of max 1% (max. 10% for cover Stone).
INNER COVER
Clean the affected area with a mild cleaning
solution for this purpose such as: alcohol with or
without tenside surfactants, Isopropanol 70 %,
oxidizing solutions, chlorine max 1 % or
hydrogen peroxide max 1,5 %.
MATTRESS
1. Disconnect the tube connector from the
control unit and remove the air from the
mattress.
2. Remove the covers.
3. Move the mattress to a clean area that is
suitable for cleaning.
4. Wipe off the air cells, all tubes and the CPR
module with a mild cleaning solution for this
purpose such as: alcohol with or without tenside
surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % or hydrogen
peroxide max 1,5 %. Let it dry.
5. Wipe off the work surfaces using a suitable
detergent and/ or disinfectant.
6. Put the mattress together. If air cells for any
reason have become loose from the tubes, these
must be replaced according to the drawing in
section 6.2.
FOAM CORE
Clean the affected area with a mild detergent
(such as washing-up liquid) and water or with an
alcohol-based disinfectant (cleaner intended for
this purpose. Gently squeeze out any water).
Note!
• Check the hygiene cover, air cells and hoses
each time the product is cleaned. If
damaged, it must be replaced or repaired.
Also check the control unit, tube connectors
and power cable during cleaning. Damaged
parts must be replaced or repaired.
Do not wring or roll the foam core to extract
the water. Let it dry in a warm, ventilated
area (not in direct sunlight). The foam core
must be completely dry before it is used
again.
17