5.
Remove the attachment for pressure compensation and the protection cap
on the USB connector. Before use, check that there is no moisture on the USB
connector.
6.
Move the deflection lever several times and check that it moves easily and that
the distal end moves properly.
7.
Check the ATMOS Scope for leak tightness using the manual leakage tester (see
chapter "3.6.2 Performing a leakage test" on page 22).
8.
Check whether an image is displayed and the LED lamp is turned on.