3.1 Continuous Wave Signal
The most common output signal of a signal generator is the continuous wave limit signal. In the following example,
we use Agilent N9020A spectrum analyzer to measure the CW output signal information of S1132 signal generator
(with frequency of 1GHz, amplitude of -10dBm).
Operation steps:
1. Instrument connection
Use a coaxial cable to connect the signal output end of the signal generator to the RF input port of the N9020A
spectrum analyzer.
2. Parameter setting
(1) Reset the signal generator
— Press【Preset】
(2) Parameter setting
— Press【FREQ】
— Press【AMPT】

3 Operation Guide

button
button, input 1 GHz
button, input -10 dBm