Figure 3-2 Test result of the frequency sweep signal
3.2.2 Frequency + Amplitude sweep signal
Frequency sweep signal refers to the frequency and amplitude sweep from the start value to the end value at the
same time.
For example: to output a sweep signal, with frequency from 1GHz to 2GHz, amplitude from -50dBm to -10dBm,
number of step points of 50, dwell time of 20ms.
Operation steps:
1. Instrument connection
Use a coaxial cable to connect the signal output end of the signal generator to the RF input port of the N9020A
spectrum analyzer.
2. Parameter setting
(1) Reset the signal generator
— Press【Preset】
button
(2) Parameter setting