3.3.2 FM
For example: frequency offset 500kHz, modulation rate 50kHz
Operation steps:
1. Instrument connection
Use a coaxial cable to connect the signal output end of the signal generator to the RF input port of the N9020A
spectrum analyzer.
2. Parameter setting
(1) Reset the signal generator
— Press【Preset】
(2) Parameter setting
— Press【FREQ】
— Press【AMPT】
— Press【RF1】
— Press【FM/ФM】
— Select [Mode] -- "FM"
— Select [Freq Offset], input 500 kHz
— Select [Int Fctn]
— Select [Freq], input 50 kHz
— Press【Return】
— Select [FM State] -- "On"
— Press【MOD】
3. Test result
After setting the relevant parameters of the spectrum analyzer, the result is shown in Figure 3-5.
button
button, input 1 GHz
button, input -10 dBm
button, make the【RF1】button backlight bright
button
button
button, make the【MOD】button backlight bright
Tel: 886. 909 602 109
17
Email: sales@salukitec.com
www.salukitec.com