The followings table shows the preset key assignment and which combinations are possible:
Function/push-
Up short
button
Menu
White balance
Light source on/off
Photo
Video
Brightness +
Brightness -
Saturation +
Saturation -
Digital zoom +
Digital zoom -
Sharpness +
Sharpness -
Contrast +
Contrast -
■ (default)
Empty
Return to factory settings
In order to return to factory settings:
Start menu (button A 20 (up), press and hold).
Mark "System setting".
Actuate button B 21 (right). A submenu appears.
Select "Return to default settings".
End application
Press the "Power off" switch 1 on the camera control unit.
Disconnect light guide cable.
Disconnect endoscope.
To completely disconnect the device from the power supply, remove the
plug from the mains connector 6 on the rear of the device.
6.

Validated reprocessing procedure

6.1

General safety instructions

Note
Adhere to national statutory regulations, national and international stan-
dards and directives, and local, clinical hygiene instructions for sterile pro-
cessing.
Note
For patients with Creutzfeldt-Jakob-disease (CJD), suspected CJD or possi-
ble variants of CJD, observe the relevant national regulations concerning
the reprocessing of products.
Note
Successful processing of this medical device can only be ensured if the pro-
cessing method is first validated. The operator/sterile processing technician
is responsible for this.
The recommended chemistry was used for validation.
Top long
Down short
■ (default
and fixed)
■ (default)
Down long
Left short
■ (default)
■ (default)
Note
If there is no final sterilization, then a virucidal disinfectant must be used.
6.2

Single-use products

WARNING
Risk of infection for patients and/or users and impairment of product
functionality due to reuse. Risk of injury, illness or death due to con-
tamination and/or impaired functionality of the product!
Do not reprocess the product!
Designation
Disposable sterile cover
6.3

General information

Dried or affixed surgical residues can make cleaning more difficult or inef-
fective and lead to corrosion. Therefore the time interval between applica-
tion and processing should not exceed 6 h; also, neither fixating pre-
cleaning temperatures >45 °C nor fixating disinfecting agents (active
ingredient: aldehydes/alcohols) should be used.
Excessive measures of neutralizing agents or basic cleaners may result in
a chemical attack and/or to fading and the laser marking becoming
unreadable visually or by machine for stainless steel.
Residues containing chlorine or chlorides e.g. in surgical residues, medi-
cines, saline solutions and in the service water used for cleaning, disinfec-
tion and sterilization will cause corrosion damage (pitting, stress corro-
sion) and result in the destruction of stainless steel products. These must
be removed by rinsing thoroughly with demineralized water and then dry-
ing.
Additional drying, if necessary.
Left long
Right short
■ (default)
■ (default)
Art. no.
JG904
en
Right long